STHV800:Octal ±90 V, ±2 A, 3-level RTZ, high-speed ultrasound pulser

The STHV800 is an octal, monolithic, high-voltage and high-speed pulse generator. It is designed for medical ultrasound applications, but can be used for other piezoelectric, capacitive or MEMS transducers.

The device integrates a controller logic interface circuit (compatible with both 1.8 V and 3.3 V input signals), level translators, MOSFET gate drivers, noise blocking diodes, and high power P-channel and N-channel MOSFETs as the output stage for each channel. These MOSFETs are capable of providing more than 2 A of peak output current. Each channel has a dedicated bridge in order to reduce power dissipation and jitter during continuous wave mode (peak current is limited to 0.3 A). This CW bridge has dedicated power supplies (HV_CW) which are fully independent on the main HV supplies.

These HV_CW supplies can be shorted to the HV supplies. The fundamental structure of each channel also consists of active clamping to ground circuitry, anti-leakage and anti-memory block, a thermal sensor to protect the device and an integrated TR-switch (just 8 Ω as equivalent resistor ) to connect the HV output to its LV output, guaranteeing strong decoupling during the transmission phase.

The eight independent T/R switches can be used in both a dedicated RX chain per channel or in a multiplexing configuration.

The clamp circuit has a current capability up to 2 A and works directly on the output pin, carrying this node exactly to zero. This feature allows minimized injection change during the transition from clamp to RX state.

In addition, the STHV800 includes self-biasing circuitry which allows very low power consumption during the RX phase (down to 200 µW global power dissipation) and thermal shutdown block sensing by an external dedicated pin (THSD).

One of the main benefits of this device is that it requires very few external components: only decoupling capacitors on the HV and LV supplies, and a resistor to pull up the THSD pin (moreover, this resistor can be shared by many devices).

Each channel is driven independently by only 2 digital bits, which in CW mode become one bit. An external clock can be used with the STHV800 to synchronize all the input signals. This feature, however, is optional: if the CK pin is tied to ground the device works in asynchronous mode.

Key Features

  • High-density ultrasound transmitter
  • Two independent half-bridges per channel
  • 0 to ±90 V output voltage
  • Power-up free
  • Synchronization of the input signals (selectable) by an external clock
  • Up to 20 MHz operating frequency
  • Low-power, high-voltage, high-speed drivers
  • 2 independently-supplied half bridges (shorted-option) for each channel, one dedicated to continuous wave (CW) mode
    • Main half bridge: ±2 A source and sink currentDown to 20 ps jitterLow 2nd harmonic distortion
    • CW half bridge:±0.3 A source and sink currentDown to 10 ps jitterVery low power consumption
  • Fully integrated real clamping-to-ground function
    • 8 Ω synchronous active clamp
    • ±2 A source and sink current
  • Fully integrated TR switch
    • 8 Ω ON resistance
    • Up to 300 MHz BW
    • Current consumption down to 10 μA in RX phase
    • Receiver multiplexing function
  • 1.8 V to 3.6 V CMOS logic interface
  • Auxiliary integrated circuits
    • Noise blocking diodes
    • Anti-leakage on output node
    • Fully self-biasing architecture
    • Thermal protection
  • Latch-up free due to HV SOI technology
  • Very few external passive components or supplies needed
产品规格
DescriptionVersionSize
DB2236: Octal ±90 V, ±2 A, 3-level RTZ, high-speed ultrasound pulser2.0630 KB
应用手册
DescriptionVersionSize
AN4471: STEVAL-IME009V1 evaluation board based on the STHV800 high voltage pulser1.11 MB
宣传册
DescriptionVersionSize
Octal ultrasound pulser High-voltage, high-speed STHV800276 KB
手册
DescriptionVersionSize
Semiconductor solutions for healthcare applications1.0665 KB
解决方案评估工具
型号制造商Description
STEVAL-IME009V1STSTHV800 Ultrasound Pulser IC evaluation board
STEVAL-IME013V1STCost effective ultrasound pulser IC evaluation board based on the STHV800
样片和购买
型号PackagePacking TypeOperating Temperature (°C) (min)Operating Temperature (°C) (max)Unit Price (US$) *QuantityECCN (EU)ECCN (US)Country of Origin
STHV800LTFLGA-56LD 8x8x0.9mmTray-40125161000NECEAR99FRANCE
质量和可靠性
型号PackageGradeRoHS Compliance GradeMaterial Declaration**
STHV800LTFLGA-56LD 8x8x0.9mmIndustrialEcopack2md_a0k7-wspc-tflga-wspc-8-wspc-x-wspc-8-wspc-x-wspc-1-wspc-56ld-wspc-pitch-wspc-0.5_acd1uj70aea-wspc-(sthv800l)_wcp_ver2_signed.pdf
md_a0k7-wspc-tflga-wspc-8-wspc-x-wspc-8-wspc-x-wspc-1-wspc-56ld-wspc-pitch-wspc-0.5_acd1uj70aea-wspc-(sthv800l)_wcp_ver2.xml
Octal ±90 V, ±2 A, 3-level RTZ, high-speed ultrasound pulser STHV800
STEVAL-IME009V1 evaluation board based on the STHV800 high voltage pulser STEVAL-IME009V1
How to upgrade the waveforms on STEVAL-IME003V1, STEVAL-IME008V1 and STEVAL-IME009V1 evaluation boards STHV748
STM8L101xx device limitations STM8L101G3
md_a0k7-wspc-tflga-wspc-8-wspc-x-wspc-8-wspc-x-wspc-1-wspc-56ld-wspc-pitch-wspc-0.5_acd1uj70aea-wspc-(sthv800l)_wcp_ver2.xml STHV800